New TopShim run:

version = topshim 2.0.0
date = 2023-10-30
time = 22:06:38.237
command = topshim  convcomp tuneaxyz

1D SHIMMING

Parameters:

maximum order = 5
probe = Z116098_0395
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 13C
o1p (from lock) = 7.24 ppm
optimisation parameters = ls
linewidth 1H = 0.10 Hz
envelope shape / strictness = 2.00 / 200

Results:

tune lock level increase = 1.3

initial B0 stdDev = 3.40 Hz
sample size = 2.84 cm, position = -0.01 cm
final B0 stdDev = 0.35 Hz > improvement = 9.7
envelope width = 0.83 Hz

Performing tune after shimming
tune lock level increase = 1.0

shim changes: 
Z	+49
Z2	-44
Z3	+142
Z4	-291
Z5	+26
X	-62
Y	+134

duration = 3 min 38 sec
completed successfully
finished Mon Oct 30 22:10:14 2023


========================================================================